Use Manual

Table Of Contents
2. Dampen a lint-free cloth with a surface disinfectant.
3. Carefully wipe the entire surface area of the monitor to remove any dust.
Related information...
Weekly cleaning procedures (IOM), page 200
Open and close the front and rear module covers (IOM), page 303
Clean the module covers (IOM)
Required materials
Antistatic plastic cleaner
Lint-free cloth
Required module
status
Offline
Perform this weekly procedure to clean the module covers.
1. Ensure that the module covers are closed and locked before the module covers are
cleaned.
2. Dampen a lint-free cloth with an antistatic plastic cleaner.
3. Wipe the entire surface area of the module cover.
4. Let the module cover air-dry to allow an antistatic film to form.
Related information...
Weekly cleaning procedures (IOM), page 200
Open and close the front and rear module covers (IOM), page 303
As-needed cleaning procedures (IOM)
As-needed cleaning procedures are required on the Input/Output Module.
Related information...
Input/Output Module cleaning, page 200
Clean the gripper (IOM), page 202
Clean the RackPorts (IOM), page 202
Clean the FlexRacks (IOM), page 203
Clean the drawers (IOM), page 203
Clean the device-specific racks of the analyzer (IOM), page 204
Clean the AccessPoint (IOM), page 204
Clean the bar code readers (IOM), page 204
Section 9
Service, maintenance, and diagnostics
Input/Output Module maintenance
GLP systems Track Operations Manual
80003968-101 - 2021-09-16
201