Input/Output Module maintenance
Before cleaning or maintenance work is performed, the module status must be transitioned to
Offline by using the Online/Offline push button.
NOTE: Follow all safety information in this manual. All work not included in this manual may only
be performed by an Abbott Laboratories representative or an authorized service representative.
Related information...
Service, maintenance, and diagnostics, page 191
Input/Output Module cleaning, page 200
Input/Output Module maintenance checks, page 205
Input/Output Module cleaning
NOTE: Do not immerse RackPorts with a radio-frequency identification (RFID) tag in water. Do
not use a dishwasher to clean RackPorts with an RFID tag.
NOTE: Do not use a dishwasher to clean FlexRacks.
Related information...
Input/Output Module maintenance, page 200
Weekly cleaning procedures (IOM), page 200
As-needed cleaning procedures (IOM), page 201
Weekly cleaning procedures (IOM)
Weekly cleaning procedures are required on the Input/Output Module.
Related information...
Input/Output Module cleaning, page 200
Clean the monitor (IOM), page 200
Clean the module covers (IOM), page 201
Clean the monitor (IOM)
Required materials
Laboratory-grade surface disinfectant
Lint-free cloth
Required module
status
Offline
Perform this weekly procedure to clean the monitor.
1. Ensure that the module covers are closed and locked before the monitor is cleaned.