Use Manual

Table Of Contents
NOTE: To avoid sample contamination, perform maintenance procedures only when no samples
are present.
1. Carefully remove any dust from the guiding slot with a cotton swab.
2. Carefully remove any dust from the guiding slot with a soft brush.
Related information...
As-needed cleaning procedures, page 193
Track maintenance checks
The following maintenance checks are required for the track to maintain optimal system
performance.
Maintenance check Interval
Verify that the lane elements are free from damage or dust. Weekly
Verify that the cross switch controllers are free from dust or foreign objects. Daily
Verify that the track hoods are closed and locked. Daily
Verify that no recaps are present on the track. Daily
Related information...
Track maintenance, page 192
Service, maintenance, and diagnostics
Track maintenance
Section 9
194
GLP systems Track Operations Manual
80003968-101 - 2021-09-16