CAUTION: Incorrectly loaded samples may constitute a health hazard for patients.
The input area can be configured in different ways, which may increase the risk of
samples being loaded incorrectly. Samples loaded in the wrong area can cause delayed
or incorrect results, and can cause highly sensitive analyzers to malfunction. Only
allow trained personnel to operate the laboratory automation system (LAS). Ensure that
samples are loaded into the defined area for the sample type (for example, centrifuged,
uncentrifuged, capped, or uncapped) on the LAS.
Figure 18: Sample tubes in FlexRacks in the module drawer
1.
To open the appropriate drawer, press the push button above the drawer (or tap the Open/
Close button on the touchscreen user interface).
IMPORTANT: Do not pull on the drawer handle to open the drawer. Pulling on the drawer
handle causes an error status for the drawer and does not fully open the drawer.
2. Load sample tubes in the correct area according to tube type, capped, uncapped, spun, or
unspun.
IMPORTANT: Samples loaded in the wrong area can cause delayed or incorrect results.
3. Verify that each sample has a bar code label.
Section 5
Operating instructions
Input/Output Module operation
GLP systems Track Operations Manual
80003968-101 - 2021-09-16
169