Use Manual

Table Of Contents
Perform this procedure to place the module offline. All processes running in the module stop.
CARs are no longer routed to the module.
1. Press the Online/Offline push button on the front of the module for a minimum of 3 seconds
or tap the green arrow area of the Online/Offline button on the touchscreen user interface.
2. Wait for the module to transition to a status of Offline.
The Online/Offline push button is illuminated steady yellow. The arrow area of the Online/Offline
button is gray.
Related information...
Input/Output Module operation, page 162
Delete an area (IOM), page 67
Pause the module (IOM)
Prerequisite The Online/Offline push button and the Online/Offline button
are illuminated steady green.
Required module
status
Online
Perform this procedure to pause the module.
The pause function is activated by using the Online/Offline push button. When the module is
paused, all processes running in the module stop. However, the connection to the Track Sample
Manager remains intact. CARs are still routed to the module.
1.
Briefly press the Online/Offline push button or tap the gray area of the Online/Offline button
on the touchscreen user interface.
2. Wait for the module to transition to a status of Pause.
The Online/Offline push button and the Online/Offline button blink green.
NOTE: If the module is paused for longer than 5 minutes, the module automatically transitions to
a status of Offline.
Related information...
Input/Output Module operation, page 162
Deactivate pause mode (IOM)
Prerequisite
The Online/Offline push button and the Online/Offline button
blink green.
Required module
status
Pause
Section 5
Operating instructions
Input/Output Module operation
GLP systems Track Operations Manual
80003968-101 - 2021-09-16
165