Use Manual

Table Of Contents
Search for samples (TSM)
Perform this procedure to search for samples on the Track Sample Manager (TSM).
1.
On the TSM main menu, tap Samples.
2.
On the Samples tab, tap the Search tab.
3.
On the Search tab, enter the TSM sample number or the sample bar code.
NOTE: The sample bar code can be entered in the following ways:
The bar code 0999.0999-9 can be entered as 0999.09999 if the exact bar code is
known.
The bar code 0999.0999-9 can be entered as 0999.0999* if the last digit of the bar
code is unknown or if there are multiple samples on the track that end with this bar
code. All samples that have a bar code beginning with 0999.0999 are listed.
4.
Tap Search or press the Enter key to display a list of search results.
5. If multiple samples are displayed for one bar code, select the sample based on its status.
NOTE: The following actions can cause multiple samples to be displayed for one bar code:
A bar code containing an asterisk was entered.
Tubes have been repeatedly removed from and placed back on the GLP systems
Track.
Additional secondary tubes with an identical bar code have been generated on the
Aliquot Module.
Related information...
Samples tab (TSM), page 121
Expert search tab (Samples) element descriptions (TSM), page 129
Remove a sample (TSM)
Perform this procedure to remove a manually removed sample from the Track Sample Manager
(TSM).
CAUTION: Do not remove samples from a CAR or the track. All samples removed from
the track must be placed back in the Input/Output Module (IOM) for appropriate routing.
When samples are removed from the track, they must be deleted from TSM before they
are reloaded.
IMPORTANT: A sample that was manually removed from the track must be marked as Removed.
Consequently, the sample receives a status of Unknown.
1.
On the TSM main menu, tap Samples.
2.
On the Samples tab, tap the Search tab.
3.
On the Search tab, enter the TSM sample number or the sample bar code.
NOTE: The sample bar code can be entered in the following ways:
Operating instructions
Track Sample Manager user interface
Section 5
128
GLP systems Track Operations Manual
80003968-101 - 2021-09-16