Use Manual

Table Of Contents
Related information...
As-needed cleaning procedures, page 29
Clean the guiding slot
Required materials
Handheld vacuum (recommended)
Extra thin cotton swab
Soft brush
Required module
status
Offline
Perform this as-needed procedure to clean the guiding slot.
NOTE: To avoid sample contamination, perform maintenance procedures only when no samples
are present.
1. Remove dust from the guiding slot with the handheld vacuum cleaner.
2. Carefully remove any dust from the guiding slot with a cotton swab.
3. Carefully remove any dust from the guiding slot with a soft brush.
Related information...
As-needed cleaning procedures, page 29
Clean the track hood
Required materials
Surface disinfectant used in the laboratory
Lint-free cloth
Required module
status
Offline
Perform this as-needed procedure to clean the track hood.
1. Ensure that the track hood is closed before it is cleaned.
2. Dampen a lint-free cloth with a surface disinfectant.
3. Carefully wipe the track hood to remove any dust.
Related information...
As-needed cleaning procedures, page 29
Clean the AccessPoint
Required materials
Surface disinfectant used in the laboratory
Service, maintenance, and diagnostics
Cleaning and maintenance Section 8
30 Sample Access Line (SAL) Interface Manual
80003970-101 DRAFT - 2022-MM-DD