Use Manual
Table Of Contents
- Table of contents
- Read me first
- Section 1: Use or function
- Section 2: Installation procedures and special requirements
- Section 3: Principles of operation
- Section 4: Performance characteristics and specifications
- Section 5: Operating instructions
- Section 6: Operational precautions and limitations
- Section 7: Hazards
- Section 8: Service, maintenance, and diagnostics
- Section 9: Troubleshooting
- Appendix A: Alinity ci-series SAL
- Design and function
- Technical data
- Sample Access Line operation
- Buttons
- Start screen
- Power on the Sample Access Line
- Place the Sample Access Line online
- Place the Sample Access Line offline
- Pause the Sample Access Line
- Deactivate pause mode
- Restart the Sample Access Line
- Main menu screen element descriptions
- Login screen
- Access the Login screen
- Information screen element descriptions
- Access the Information screen
- Access the Configuration screen
- Error message screen
- Acknowledge an error message
- Acknowledge a message
- Weekly cleaning procedure
- Error messages
- Communication and software update error messages (0-10000)
- Module component error messages (20000-29999)
- Message code: 20105
- Message code: 20200
- Message code: 20201
- Message code: 20202
- Message code: 20203
- Message code: 20204
- Message code: 20205
- Message code: 20206
- Message code: 20207
- Message code: 20209
- Message code: 20210
- Message code: 20211
- Message code: 20901
- Message code: 20902
- Message code: 20910
- Message code: 20911
- Message code: 20912
- Archive error messages (21300-21468)
- Message code: 22501
- Message code: 22502
- Message code: 22550
- Message code: 22600
- Message code: 22601
- Message code: 22602
- Message code: 22603
- Message code: 22604
- Message code: 22605
- Message code: 22606
- Message code: 22607
- Message code: 22608
- Display error messages (23000-23017)
- Message code: 27000
- Error message on the touchscreen user interface
- CAR with sample at AccessPoint, in traffic jam, or in long queue; sample matter has been removed
- Revision history
- Index
Maintenance check Activity Interval
Verify that no foreign objects are present
on the track.
Remove any foreign objects. Daily
Verify that the track hoods are closed
and locked.
Close the track hoods if applicable. Daily
Related information...
Cleaning and maintenance, page 28
As-needed cleaning procedures
As-needed cleaning procedures are required on the Sample Access Line.
CAUTION: Risk of injury due to inattentiveness while opening and closing track hoods.
The operator may be injured while opening and closing the track hoods. Carefully open
and close the track hoods. Ensure that no body parts are placed between the housing
and the track hoods.
Related information...
Cleaning and maintenance, page 28
Clean the lane elements, page 29
Clean the guiding slot, page 30
Clean the track hood, page 30
Clean the AccessPoint, page 30
Clean the barcode reader, page 31
Clean the lane elements
Required materials
• Handheld vacuum (recommended)
• Surface disinfectant used in the laboratory
• Lint-free cloth
Required module
status
Offline
Perform this as-needed procedure to clean the lane elements.
NOTE: To avoid sample contamination, perform maintenance procedures only when no samples
are present.
1. Remove dust from the lane elements with the handheld vacuum cleaner.
2. Dampen a lint-free cloth with a surface disinfectant.
3. Carefully wipe the lane elements to remove any dust.
Section 8
Service, maintenance, and diagnostics
Cleaning and maintenance
Sample Access Line (SAL) Interface Manual