I-E67-84-1B February,2002 8-13
4 describes the function and programming mode
of each state.
Table 8-4. Temperature Compensation States
State Function Programming
Mode
MANUAL Used when a fixed temperature value Basic/Advanced
can be applied instead of a measured
value. The initial value is set at
25 C. Use the Temperature Calibrate
o
State to change the fixed temperature
value. Nernstian compensation is
applied using the fixed temperature
value.
AUTO Used when a measured temperature value Basic/Advanced
is being provided by a temperature
sensor. Nernstian compensation is
applied using the measured value.
AUT.SOL Used when a measured temperature value Advanced
is being provided by a temperature
sensor. Nernstian compensation and a
solution coefficient is applied using
the measured value.
Manual Nernstian State (Basic/Advanced)
Manual temperature compensation adjusts for
temperature effects of the sensor at a
specific temperature. A temperature sensor is
not required (i.e., NONE for TMP.SNS). The
Manual Nernstian State is only applicable for
pH Analyzer States. The displayed process
variable (i.e., pH) is standardized to 25 C.
o