User guide
I-E67-84-1B February,2002 8-5
Table 8-1. Modify Configure States
State Function Programming
Mode
ANALZER Used to define the type of analyzer. Choices Basic/Advanced
include pH, ORP, pION, and Ion Concentration
(ION.CON).
TMP.SNS Used to define the type of temperature sensor. Basic/Advanced
Choices include None, Pt100, and 3k Balco.
TC.TYPE Used to define the type of temperature Basic/Advanced
compensation. Choices include Manual Nernstian,
Automatic Nernstian, and Automatic Nernstian
with Solution Coefficient.
AO1.OUT Used to set Analog Output One range. Basic/Advanced
AO2.OUT Used to set Analog Output Two range. Basic/Advanced
RELAY1 Used to set Relay Output One function and Basic/Advanced
parameters. Choices include Setpoint, Cycle
Timer, Diagnostics, and Cleaner.
RELAY2 Used to set Relay Output Two function and Basic/Advanced
parameters. Choices include Setpoint, Cycle
Timer, Diagnostics, and Cleaner.
RELAY3 Used to set Relay Output Three function and Basic/Advanced
parameters. Choices include Setpoint, Cycle
Timer, Diagnostics, and Cleaner.
DAMPNG Used to reduce fluctuation in the display values Basic/Advanced
and output signals.
DIAGS Used to set the sensor diagnostics ON or OFF. Basic/Advanced
SAF.MD.1 Used to define the output signal state for Basic/Advanced
Analog Output One when a detected error results
in a condition that renders the analyzer
inoperable. Choices include fail Low or fail
High.
SAF.MD.2 Used to define the output signal state for Basic/Advanced
Analog Output Two when a detected error results
in a condition that renders the analyzer
inoperable. Choices include fail Low or fail
High.
SPIKE Used to set the spike magnitude level. Advanced
As with the other modes and states of
operation, the NEXT Smart Key provides access
to all Modify Configure States. Pressing the
NEXT Smart Key sequentially moves the user
through each state. This cycle repeats until
a Modify Configure State is selected using the
SELECT Smart Key or the escape function is
chosen using the Exit To MEASURE Smart Key.
Use Figure 8-2 and 8-3 to identify the Smart
Key assignments and the resulting action.
When selecting a Modify Configure State, the
configured (i.e., active) item within that
state will be the first item shown. This item
will remain the configured item until a new
item is entered and the configuration saved.