User guide

I-E67-84-1B February,2002 8-3
pressing the ENTER Smart Key when viewing the
PT 100 option. An intermediate confirmation
screen will query the user on their desire to
modify this option using the YES and NO Smart
Keys. If the Modify Configure State has been
secured, the security code will be requested.
Upon entering the correct code or if the
Modify Configure State has not been secured,
the TB84PH Advantage Series analyzer will go
directly to TMP.SNS Modify Configure State and
allow the user to change the temperature
sensor type. After completing the change,
pressing Exit To MEASURE Smart Key moves the
user to the configuration SAVE? State.
Pressing the YES Smart Key saves the new
temperature sensor option and returns the
analyzer to the Measure Mode.
BASIC/ADVANCED PROGRAMMING MODE
The Configure Mode is split into two
programming groups: Basic and Advanced. These
two options are specified by nomenclature and
control the number of configuration options
available in the Modify Configure State.
The Basic Programming Mode contains a subset
of configuration options found in the Advanced
mode. Separation into two programming groups
is advantageous when limited functionality is
desired. Fewer options reduces confusion and
the possibility of configuration errors.
When Advanced programming is ordered, the
programming toggle (i.e., Basic/Advanced) must
be set in two locations: the User State in the
Utility Mode and the Modify Configure State in
the Configure Mode. In order to select either
the Basic or Advanced Programming Mode in the
Modify Configure State, the Programming Mode
must be set to Advanced in the User State.
See Section 11, Utility Mode, for more
information on setting the User State
programming mode to Advanced.
When in the Configure Mode and Advanced
programming has been set in the User State,
the TB84PH Advantage Series analyzer queries
if the user would like Basic programming.
Pressing the ENTER Smart Key moves the user to
the Modify Configure States, pressing the NEXT
Smart Key moves the user to the Advanced
programming query, and pressing the Exit To
MEASURE Smart Key escapes to the Measure Mode.
To set the analyzer to Advanced programming,
the user presses the ENTER Smart Key when
queried to set the programming to Advanced.