User guide
I-E67-84-1B February,2002 8-1
SECTION 8 - CONFIGURE MODE
INTRODUCTION
The Configure Mode of Operation establishes
the operating parameters of the TB84PH
Advantage Series analyzer. These parameters
include programming type, analyzer type,
sensor type, temperature compensation type,
analog output ranges, relay output parameters,
damping value(s), diagnostic functionality,
safe mode levels, and spike magnitude (i.e.,
level).
A description of each configuration item and
related parameters will be included. Review
each of the following sections before
configuring the TB84PH Advantage Series
analyzer.
PRECONFIGURATION DATA REQUIRED
Before attempting to configure the TB84PH
Advantage Series analyzer, the following
requirements must be defined.
1. Analyzer parameters.
2. Analog Output Range values.
3. Relay Output function and parameters.
4. Security requirements.
5. Sensor Diagnostic functionality.
Use the worksheets found in Appendix C to help
establish the proper settings for any given
application. Use these sheets during the
configuration entry procedure and retain them
as a historical record for future reference.
CONFIGURE VIEW/MODIFY STATE
Upon selecting the Configure Mode of
Operation, a decision point is reached to
Modify or View the configuration of the TB84PH
Advantage Series analyzer. The Modify
Configure State enables analyzer options to be
set and saved into memory. In order to
provide the ability to secure the Modify
Configure State yet leave the ability to view
configuration information, the View Configure
State can be entered without using a security
code.