User guide
I-E67-84-1B February,2002 7-10
Spike State
The Spike State toggles the operational state
of the spike output function. The spike
function modulates the current output on
Analog Output One by the amount established in
the analyzer configuration. See Section 2,
Analyzer Functionality And Operator Interface
Controls, and Section 8, Configure Mode, for
more information.
Toggle the Spike output using the following
procedure:
1) Select the Spike State of Operation using
the SELECT Key.
2) Toggle the spike output function to the
desired state (i.e., OFF or ON) using the
Key, and press the ENTER Key to
accept. Press the Exit To MEASURE Key to
escape to the Measure Mode.
Note: Once the Spike State is OFF, changing the
configured spike level in the Configure Mode
will not re-enable the Spike State. The Spike
State can only be turned ON or OFF in the
Output/Hold Mode of Operation.