I-E67-84-1B February,2002 7-7
(determined by the analyzer’s
configuration) using the Key to
increment the blinking digit and the
Key to move to the next digit and press
the ENTER Key to enter the new value, or
press the ENTER Key to continue to the 20
milliampere value. Press the Exit To
MEASURE Key to escape to the Measure
Mode.
3) Press the ENTER or Exit To MEASURE Key to
escape to the Measure Mode, or edit the
process variable value for the 20
milliampere point using the Key to
increment the blinking digit and the
Key to move to the next digit and press
the ENTER Key to enter the new value.
Note: If 1)the zero or four milliampere value is
changed, 2)the new value is valid per the
specification in Table 1-3, 3)this change is
accepted using the Enter Key, and 4)the user
escapes to the Measure Mode using the Exit To
Measure Key without adjusting the 20 milliampere
value, the output range will now reflect the
newly entered zero or four milliampere point.
Analog Output Two Rerange State
The Analog Output Two Rerange Output/Hold
State provides the ability to change the
output range of Analog Output Two. One or
both end point values can be changed to any
value or range of values that are within the
specifications listed in Table 1-3.
Conduct a Rerange of the output values using
the following procedure:
1) Select the Rerange State of Operation
using the SELECT Key.