User guide

I-E67-84-1B February,2002 6-11
value using the Key to increment the
blinking digit and the Key to move to
the next digit. Press the ENTER Key to
enter the new value or to proceed to the
sensor offset value. Press the Exit To
MEASURE Key to escape to the Measure
Mode. Valid slope values range from 40
to 150%.
3) Edit the sensor PV OFF (i.e., offset)
value using the Key to increment the
blinking digit and the Key to move to
the next digit. Press the ENTER Key to
enter the new value or to proceed to the
temperature slope value. Press the Exit
To MEASURE Key to escape to the Measure
Mode. Valid offset values range from
-1000 to +1000 millivolts.
4) Edit the temperature TMP.SLP value using
the Key to increment the blinking
digit and the Key to move to the next
digit. Press the ENTER Key to enter the
new value or to proceed to the
temperature offset value. Press the Exit
To MEASURE Key to escape to the Measure
Mode. Valid slope values range from 0.2
to 1.5.
5) Edit the temperature TMP.OFF value using
the Key to increment the blinking
digit and the Key to move to the next
digit. Press the ENTER Key to enter the
new value or to proceed to the Edit
Calibrate State. Press the Exit To
MEASURE Key to escape to the Measure
Mode. Valid offset values range from -40
to +40 C.
o