User guide
TMPCAL
NO
023
MEASURE YES
MEASURE
STABL?
023
ENTER
NEW.VAL
MEASURE
TMP.CAL
RETURN
ENTER
UNITS
MEASURE
C
o
Toggle between degrees C and degrees F.
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SPT/TUNE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SPT/TUNE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SPT/TUNE
If calibration succeeds, store
data and return to "TMP.CAL"
If calibration fails, display "BAD.CAL"
and do not save data.
C
O
C
O
Show last selected units.
I-E67-84-1B February,2002 6-9
Figure 6-4. Screen Flow Diagram For
Temperature Calibrate State of Operation.
Conduct a Temperature Calibration using Figure
6-3 and the following procedure:
1) Before installing the sensor into its
final installed location, allow the
sensor to reach ambient temperature.
2) Select the Temperature Calibrate State of
Operation using the SELECT Key.