User guide

I-E67-84-1B February,2002 3-16
4) Connect the BNC Adapter pin terminals to
the corresponding SENSOR CONNECTIONS terminal
block locations via the conductor color
coding.
Note: If the sensor does not have a female
connector for the Temperature Compensator
(TC), connect the leads from the sensor
directly to the terminal block locations TB2-5
and TB2-6 (i.e., Red and White). For sensors
without TC, leave TB2-5 and TB2-6 open.
Figure 3-8. BNC Adapter Installation and Wiring Diagram
5) Connect the male BNC from the sensor to the
female BNC from the BNC Adapter and the female
TC connector from the sensor to the male TC
connector from the BNC Adapter.
6) Slide the protective boot over the BNC
connection to prevent any possible shorting to
earth surfaces.
Note, if the BNC surface is in contact with a
earth ground surface, a ground loop through
the sensor may occur and could result in poor
sensor performance and shortened sensor life.