User guide
(L2)
(L1)
REAR VIEW
SENSOR CABLE TO
BE SEALED IN
INTERNAL GROUND
TERMINALS
EXTERNAL
GROUND
TERMINAL
LINE
NEUTRAL
EARTH
INPUT
POWER
CONTROLPROCESS
RECORDER DEVICE
ANNUNCIATOR
ANNUNCIATOR ANNUNCIATOR
1
3
4
5
6
7
8
BLUE
YELLOW
BLACK
GREEN
RED
WHITE
HVY GRN
2
SENSOR CONNECTIONS
TB2
POWER/OUTPUT
CONNECTIONS
TB3
2175 LOCKHEED WAY
CARSON CITY, NV 89706
SENSE
GUARD
REF
SOL GND
RTD
RTD
SHIELD
1
3
4
5
6
7
8
2
1(+)
1(-)
2(+)
2(-)
1
2
9
10
3
3
1
3
2
LINE
CHASSIS GROUND
TB1
2
1
NEUTRAL
I-E67-84-1B February,2002 3-14
permanently damage the sensor.
Figure 3-7. Instrument Wiring Diagram.
Non-Advantage Series (TB5) Sensor Wiring
For standard and Next Step ABB sensors (i.e.,
TB5), a pin terminal option is available for
installation to a TB84PH Advantage Series
analyzer. Though this cable option is
recommended when using a non-Advantage Series
sensor, stocking preferences may tend towards
the use of one common sensor. If this sensor
type requires a BNC connector to mate with
existing instrumentation, a BNC Adapter
(included with all TB84PH Advantage Series
analyzers) provides such a connection.
When using the pin terminal sensor cable
option (i.e., the T[][] nomenclature option
where [][] represents the cable length),
connect color coded leads to the analyzer’s
terminal block as follows:
Terminal Block Sensor Color Function
Location Code
TB2-1 Blue Sense
TB2-2 N/A Guard
TB2-3 Black Reference
TB2-4 N/A Solution Ground
TB2-5 Red (If RTD
Applicable)
*