An Offset value of less than -180.0 mV or
greater than +180 mV also indicates a
potentially bad process calibration or poor
performing sensor. Calibration values that
yield Offset values less than -1000 mV or
greater than +1000 mV are not accepted.
Again, a bad calibration will be reported, and
the user returned to the beginning of the
calibration cycle.
For temperature, a bad calibration will be
reported and calibration values will not be
accepted for Slope values that are less than
0.2 or greater than 1.5 and Offset values that
are less than -40 C or greater than +40 C.
oo
Temperature calibrations use smart software
routines that automatically adjust the Slope,
Offset, or both values based on the
calibration value being entered and
calibration history if it exists.