User guide

I-E67-84-1B February,2002 12-6
ALL) text.
4) Press the SELECT Key to reset all
analyzer parameters.
5) Enter the security password 255 using
the Key to increment the blinking
digit and the Key to move to the next
digit, and press the ENTER Key to enter
the password.
6) Confirm the reset operation when the text
RESET? is displayed by pressing the YES
Key, or abort the reset operation by
pressing the NO Key.
Soft Boot User State
The Soft Boot User State initiates a software
reset. The software reset initiates boot-up
and self-test procedures. All programmable
instrument parameters are unaffected by this
function.
To reboot the analyzer without affecting any
instrument parameters, use the following
procedure:
1) Access the User Mode by pressing the
hidden button located directly above the
NT in the ADVANTAGE text on the keypad.
The text USER will appear in the
secondary display once the hidden button
has been pressed.
2) Press the SELECT Key to access the User
Mode. The text MODE will appear in the
secondary display.
3) Press the NEXT Key until the secondary
display region shows RST.SFT (i.e., Reset
ALL) text.