User guide
I-E67-84-1B February,2002 12-5
To remove the security, use the following
procedure:
1) Access the User Mode by pressing the
hidden button located directly above the
NT in the ADVANTAGE text on the keypad.
The text USER will appear in the
secondary display once the hidden button
has been pressed.
2) Press the SELECT Key to access the User
Mode. The text MODE will appear in the
secondary display.
3) Press the NEXT Key until the secondary
display region shows RST.SEC (i.e., Reset
Security) text.
4) Press the SELECT Key to reset the
security.
5) Enter the security password 732 using
the Key to increment the blinking
digit and the Key to move to the next
digit, and press the ENTER Key to enter
the password.
6) Confirm the reset operation when the text
RESET? is displayed by pressing the YES
Key, or abort the reset operation by
pressing the NO Key.
Reset All User State
The Reset All User State returns all analyzer
parameters back to factory defaults. This
includes calibration, output/hold,
configuration, security, secondary display,
and setpoint/tune values.
To reset all analyzer parameters, use the
following procedure:
1) Access the User Mode by pressing the
hidden button located directly above the
NT in the ADVANTAGE text on the keypad.
The text USER will appear in the
secondary display once the hidden button
has been pressed.
2) Press the SELECT Key to access the User
Mode. The text MODE will appear in the
secondary display.
3) Press the NEXT Key until the secondary
display region shows RST.ALL (i.e., Reset