User guide
I-E67-84-1B February,2002 12-2
Table 12-1. User States
State Function
MODE Sets the Programming Mode that are
available in the Modify Configure
State of Operation.
RST.CON Resets the configuration to factory
default settings.
RST.SEC Resets the security password and
removes all security.
RST.ALL Resets all programming parameters
such as configuration, calibration,
output/hold, security, secondary
display, and setpoint/tune
functions to factory default
settings.
RST.SFT Resets the analyzer by repeating
the boot-up and self-test
procedures.
Figure 12-1 identifies the Smart Key
assignments and resulting action. The
following section describes each of the User
States and their applicability.
Advanced/Basic Programming Mode User State
In order to simplify the configuration process
for a user who only needs a limited amount of
functionality, the TB84PH Advantage Series
analyzer contains two types of Programming
Modes: Basic and Advanced. The Programming
Mode is defined by a nomenclature option.
The Basic Programming Mode contains a reduced
set of features found in the Advanced
Programming Mode. Reducing the available
features helps streamline the configuration
process. If the TB84PH Advantage Series
analyzer is ordered with Advanced Programming,
the Basic or Advanced Programming Mode can be
used.
Contact ABB for information on Advanced
Programming upgrades.