User guide

I-E67-84-1B February,2002 11-7
Set the Cleaner parameters of a Relay Output
using Figure 8-14 as reference and the
following procedure:
1) Set the Cycle Time in hours (i.e.,
CYC.HRS) using the Key to increment
the blinking digit and the Key to move
to the next digit, and press the ENTER
Key to enter the new value.
2) Set the On Time in minutes (i.e., ON.MIN)
using the Key to increment the
blinking digit and the Key to move to
the next digit, and press the ENTER Key
to enter the new value.
3) Set the Recovery Time in minutes (i.e.,
RCV.MIN) using the Key to increment
the blinking digit and the Key to move
to the next digit, and press the ENTER
Key to enter the new value.
4) Hold the Analog Outputs (i.e., AO.HLD)
during the On and Recovery Times using
the YES Key, or leave the Analog Outputs
live during the On and Recovery Times
using the No Key.
5) Hold the Relay Outputs (i.e., RO.HLD)
during the On and Recovery Times using
the YES Key, or continue onto the Disable
Relay Outputs State during the On and
Recovery Times using the No Key.
6) Disable the Relay Outputs (i.e., RO.HLD)
during the On and Recovery Times using
the YES Key, or leave the Relay Outputs
live during the On and Recovery Times
using the No Key.