User guide
CYCLE
TIMER1
10.0
ENTER
LO SPT
MEASURE
CONFIG
EXIT
RELAY1
RETURN
ENTER
CYC.MIN
MEASURE
0.5
ENTER
ON.MIN
MEASURE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SPT/TUNE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SPT/TUNE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
SPT/TUNE
Display "HI" or "LO",
as configured.
pH
1.0
*99.9 minutes is the maximum
allowable time limit.
*
*
I-E67-84-1B February,2002 8-30
the Cycle Time. Thus, a Cycle Timer will
energize the Relay Output for a set amount of
time (ON.TIME) and de-energize for the
remainder of the cycle (CYC.MIN). This cycle
repeats until the Setpoint condition is no
longer met. For more information on the Cycle
Timer, see Section 2, Overview.
Valid Setpoint values are limited to the
Process Variable range of the TB84PH Advantage
Series analyzer. See Table 1-3,
Specifications, for analyzer range values.
Valid Cycle Time and On Time values are 0.0 to
99.9 minutes.
Figure 8-13. Screen Flow Diagram For Setting
the Cycle Timer Relay Output Configure State
of Operation.
Set the Cycle Timer parameters of a Relay
Output using Figure 8-13 as reference and the
following procedure:
1) Set the Setpoint (i.e., LO SPT or HI SPT)
using the Key to increment the
blinking digit and the Key to move to