Specifications
1. OPERATION
Item n°. : W1404621 Page 1-4 Version 01.02
1.2. Installation
MAGSCREEN is connected to MAGLIFE C PLUS. by means of a bi-directional optical fibre
cable connected to (13) on the rear of the device.
The cable requires a passage through the Faraday cage called a "wave guide", which is
normally designed on the filter plate of the MRI system or close to the HF window.
If this is not so, have a "wave guide" installed by the manufacturer of the Faraday cage in
compliance with the specifications of the MRI equipment manufacturer.
Internal free access: 17 mm minimum for a straight "wave guide" and 55 mm if there is a 90°
bend.
(the maximum outer diameter of the optical fibre connector is 16 mm)
Standard length of the optical fibre link: 25 m
MAGSCREEN has an external power supply unit which must be connected to a standard
mains power supply socket.
Mains power supply:
Voltage: 115 V (± 10 %) or 230 V (± 10 %)
Frequency: 50 Hz or 60 Hz
Power: 60 VA maximum
G
Important: MAGSCREEN is not compatible with magnetic resonance (it could
be attracted by the stray MRI magnetic field and produce artefacts
on the MR images)
consequently:
DO NOT INTRODUCE MAGSCREEN INTO THE MAGNET ROOM!