Specifications
7. PREVENTIVE MAINTENANCE
Item n° : W1404621 Page 7-1 Version July 1998
7.1. Preventive maintenance kit
7.2. Cleaning the MAGSCREEN
Clean the unit with common cleaning and disinfecting agents (BURATON, INCIDIN. GG.
KORSOLIN or LYSO FORMIN 2000).
Apply the solution on a cloth, and not directly on the device. Do not use abrasive cleaning
agents or organic solvents.
I
Important: Make sure the unit is off and disconnected from the mains
before cleaning.
Do not remove the covers.
If any liquid enters the device, have the device cleaned and
verified thoroughly.
To avoid scratching the display screen on the front of the device, clean it by blowing away dust
particles or by wiping it with a sponge moistened with a cleaning agent. Do not use abrasive
cleaning materials. Fingerprints and stains may be removed with lens cleaning solution and a
soft cloth. Do not rub the screen with a dry cloth or use hydrochloric solvents or alcohol.
The unit shall not be exposed to temperatures above 50 °C.
Do not autoclave.
Device and accessories must be decontaminated (risk of pathological genes) before being
scrapped.
7.3. Visual/mechanical/physical inspection
Perform such inspections regularly, once every 6 months.
Check unit for any obvious signs of physical damage, and replace as required.
Before each use, check all unit functions, simulate an alarm and check if all cable connectors
and insulators are in a good condition.
Once a year check the leakage current as per paragraph 19 of standard CEI 601-1.