1. OPERATION
Item n°. : W1404621 Page 1-7 Version 01.02
1.4. Transfer of control to MAGSCREEN
MAGSCREEN is used for remote controlling the operation of MAGLIFE C PLUS. To transfer
control, press rotating button (3). The following message will be displayed on the screen:
MAGSCREEN Version XX.XX
Press the Freeze key and the rotating button
simultaneously to control MAGLIFE C PLUS.
Now press Freeze key (8) while keeping rotating button (3) pressed down.
The transfer of control to MAGSCREEN is displayed by a change in the icons in the status
bars of MAGSCREEN and MAGLIFE C PLUS..
The main controls are now accessible from rotating button (3) and key (2) via menus. The
procedures are identical to those used with MAGLIFE C PLUS. (see sections 4.5. and 4.6. of
the MAGLIFE User's Manual, item code W1403956).
Keys (5) to (8) are used as with MAGLIFE C PLUS.
F
Notes:
- MAGLIFE C PLUS. keeps control of the system at all times.
Any action by the operator on the front panel of MAGLIFE C
PLUS. (keys, buttons) will unconditionally interrupt any
remote control established previously.
- remote control will be rejected or ignored by MAGLIFE C
PLUS.:
* in the following specific operating modes: "Multigas
Test", "NIBP Test" and "Option Configuration". In these
three modes, only the MAGLIFE C PLUS. controls are
effective.
* if a MAGLIFE C PLUS. control (key or rotating button)
has been used during the previous 15 seconds.
1.5. Transfer of control to MAGLIFE C PLUS.
Control is transferred back to MAGLIFE C PLUS. in the following four cases:
- 5 minutes after the last MAGSCREEN command ,
- after a command from MAGLIFE C PLUS.,
- each time MAGLIFE C PLUS. is turned on,
- each time connection with MAGLIFE C PLUS. is lost or faulty.