Product specifications

AD-4329A Weighing Indicator Page 43
11.1.1. Data Format
The data format of this SER.OUT is the same as the format 1 of the RS-232C
interface.
Refer to "12.1.2. Data Fromat" of RS-232C interface at page 45, if you need details.
Header 1
ST Stable weighing data.
US Unstable weighing data.
OL Overload (Out of range).
Header 2
GS or G~ Gross data.
NT or N~ Net data
TR or T~ Tare data.
PT Preset tare data.
Data
The first of the data bits is the polarity whether "+" or "-".
When data is zero, the polarity is "+".
Format is 8 figures (bits) including polarity and decimal point (dot or comma).
In case of "out of range", the data bits of format 1 are replaced by spaces except the
decimal point.
In case of "out of range", the data bits of the format 2 are replaced by all 9's
Unit
Function Cf-10 is used in common with RS-232C and standard serial data output.
In case of Cf-10 0 or 2, the unit length is 2 figures.
In case of Cf-10 1, or 3, the unit length is 3 figures. Depending on circumstances,
an A&D printer may not work correctly.
11. Standard Serial Data Output