Product specifications

AD-4329A Weighing Indicator Page 35
8.1.2. Display and Operation
Action of Accumulating Data
When accumulating data, the display blinks once.
If storing accumulation data, the M+ mark is displayed.
Caution This function can not accumulate different unit data. Specify a unit before use.
Display of Accumulation Data
When setting to Cf-11 1 (Effective accumulation function) and pressing the
MODE key, the total is displayed and the total data is displayed with the M+
mark blinking . If the MODE key is pressed again, the weighing data is displayed.
The total data can be output. Refer to "Output of Accumulation Data"
If "lb/kg" of Cf-01 3 is selected, the display of "lb", "kg" and "total" is exchanged
with the MODE key in order.
Undoing the Accumulation Data
The last weighing data can be deduced from the accumulation data unless new
data is accumulated.
Step 1 Press the MODE key to display total and accumulation data.
Step 2 Press and hold the M+ key more than 3 seconds. The display blinks once and
the last accumulation data before accumlating last weighing data is displayed.
Caution External input can not be used.
Clearing the Accumulation Data
Step 1 Press the MODE key to display total and accumulation data.
Step 2 Press and hold the ZERO key more than 3 seconds. The display blinks once
and the accumulation data is cleared.
Caution External input can not be used.
Output of Accumulation Data
Accumulation data can be output to the standard serial output and RS-232C
interface using the manual print mode.
Step 1 Press the MODE key to display total and accumulation data.
Step 2 Pressing the PRINT key, the count and total of accumulation data are output in order.
Example:
8. Accumulation