Product specifications
AD-4329A Weighing Indicator Page 33
7. Tare
The function is used to display a net value with the container weight subtracted from
the total weight, if you put an object into a container to weigh it.
There are two methods for the "weighing tare weight" and "digital input of tare
weight".
Using the RS-232C interface, you can perform digital input from external equipment.
Caution When turning the display off with Cf-05 1, the tare value is cleared.
When turning the power off, the tare value is cleared.
7.1.1. Weighing Tare
Operation Put the tare on the weighing unit. Press the TARE key to store the tare
weight after the MD mark turns off. The dispaly changes to net.
Caution When displaying a negative gross value, tare can not be used.
7.1.2. Digital Input
Caution The input value is rounded off to the unit of division (weighing interval).
In case of Cf-05 1(power-on zero), the display value will be zeroed
when turning the display on.
When using a multi-interval scale, usable input range is the first range.
In case of Cf-06 1(To inhibit preset tare), preset tare can not be used.
Step 1 Press the
PRESET TARE
key to display the stored tare value.
When tare is cleared or is not used, the value is zero.
Step 2 Enter a new tare value using the following keys.
key...........The key to select a figure.
key...........The key to select a number.
key...........The key to store a new tare value.
The net is displayed.
7.1.3. Clearing Tare
When pressing the TARE key while gross is zero, tare is cleared and gross is
displayed.
When zeroing with the ZERO key, tare is cleared.
7. Tare