Product specifications
AD-4329A Weighing Indicator Page 31
Description of "Stream Mode"
Object f-31 0 , f-41 0
Operation Data is output in every sampling (when refreshing the display).
Use this mode to output data to an external display (Data may not be output
due to timing of the baud rate and internal sampling rate). If data is printed
with pressing the PRINT key on the printer, use the stream mode.
Description of "Auto Print Mode"
Object f-31 1 , f-41 1
Operation When weighing data varies from the "inhibition region of output" to the
"permission region of output", the stable data is just output once. If you use
this mode, set f-02 except 0.
When weighing (and removing) each object and printing the data, use this
mode.
In case of f-32 0 or, f-42 0 .
"Inhibition region of output" +5d.
+5d < "permission region of output".
In case of f-32 1 or, f-42 1 .
-5d ≤ "inhibition region of output" +5d.
"Permission region of output" < -5d, +5d < "permission region of output".
d: division (weighing interval) of first range.
Description of "Manual Print"
Object f-31 2 , f-41 2
Operation When pressing the PRINT key, the stable weighing data is just output once
Description of "Delay for continual data"
Object f-33, f-43
Operation This function can be used in the "Auto print mode" and "Manual print mode"
When using a non-buffered printer like the AD-8121, set to f-33 1 and
f-43 1 .
6. Functions