Product specifications
AD-4329A Weighing Indicator Page 27
6.2. F-Functions
Digital Filter
Item
Parameter
Description
0
2 d/ 1.6s
1
4 d/ 1.6s
2
8 d/ 1.6s
3
16 d/ 1.6s
4
32 d/ 1.6s
5
64 d/ 1.6s
6
128 d/ 1.6s
7
2 d/ 3.2s
* 8 4 d/ 3.2s
9
8 d/ 3.2s
10
16 d/ 3.2s
11
32 d/ 3.2s
12
64 d/ 3.2s
f-00
Filter
Motion / Averaging
time
13
128 d/ 3.2s
If weak filter is set, the
response will be fast, but will
be more sensitive to external
influences such as vibration.
0
OFF
1
0.5 d/ 1s
2
1.0 d/ 1s
3
1.5 d/ 1s
4
2.0 d/ 1s
5
2.5 d/ 1s
6
0.5 d/ 2s
7
1.0 d/ 2s
* 8 1.5 d/ 2s
9
2.0 d/ 2s
f-01
Zero tracking
10
2.5 d/ 2s
This function traces the
weighing value drifting around
zero point slowly, displayed as
zero. If a strong parameter is
set, a very small zero drift may
be not detected.
If Cf-03 is 1, the 0, 1, 6 or 7
can be selected only.
If Cf-03 is 1, the initial setting
sets 7.
0
No motion detection
1
0.5 d/ 0.5s
2
1.0 d/ 0.5s
3
2.0 d/ 0.5s
4
3.0 d/ 0.5s
5
4.0 d/ 0.5s
6
0.5 d/ 1s
7
1.0 d/ 1s
* 8 2.0 d/ 1s
9
3.0 d/ 1s
f-02
MD mark
Motion detection
condition
10
4.0 d/ 1s
The function to set the
condition of judgment whether
a weighing value is unstable or
stable. The ZERO key and
TARE key has effective in the
stable state. If these keys need
in the unstable state, set to
f-02 0. In case of Cf-03 1,
f-02 6 or f-02 7 can be
selected only. (Internal setting
is 6.)
d: division (weighing interval) of first range.
s: second.
*: factory settings (Initial settings).
6. Functions