5.6. Calibration Error Code List
Exiting from the calibration error
MODE key .. The key to return the point occurred an error. Retry the operation.
Press and hold the
NET/GROSS
key and press the CAL key
All parameters are not changed, the CanCel is displayed and the
calibration mode is finished.
Error Code List
If an error has occurred during calibration mode, the following code is displayed.
Error code
Description
err 0
In multi-interval scale. The last division is set to maximum ( d-50 ).
Therefore the next division can not be entered.
err 1
Resolution exceeds 10000. (Resolution = maximum capacity/ minimum
division)
Reduce maximum capacity or increase minimum division.
err 2
err 3
Load cell output is too large or too small at zero calibration. Check the weighing
unit and load cell. Refer to "3.1.4.Verifying Load Cell Output and Input
Sensitivity".
err 4
Measuring calibration mass (or entering digital input), the value exceeded
maximum capacity. Reduce calibration mass (or digital input).
err 5
Selected calibration mass is smaller than minimum division.
err 6
New input sensitivity is less than 0.15 µV/division. Increase input sensitivity.
Refer to "3.1.4.Verifying Load Cell Output and Input Sensitivity".
err 7
Placing mass on the weighing unit, the load cell output becomes a negative
value. Check the load cell cable connections and the direction of load cell
mounting.
err 8
Load cell output exceeds the input range before maximum capacity.
Adjust zero balance referring to "3.1.4.Verifying Load Cell Output and Input
Sensitivity". Replace with a load cell designed smaller output. Reduce
maximum capacity.
err 9
The weighed value is out of the input range at zero calibration or span
calibration. Check the weighing unit and cables.
err 12
The last weighing range is larger than next weighing range.
err 13
Incorrect mass is selected at the digital linearization function.
Select mass of the following relation. lnr 1 < lnr 2 < lnr 3.
g err
An unacceptable value was selected in the gravity acceleration function.
Exit this error using the
Operate/Standby
key.
5.6. Calibration Error Code List