Product specifications
AD-4329A Weighing Indicator Page 23
5.5. Gravity Compensation Function
If the scale is used at the calibration location, it is not necessary to perform this function.
It may cause a weighing error, if there is a difference of gravity acceleration
between the installed location and calibration location. This function specifies
these gravity accelerations and corrects span error.
Caution The decimal point is not displayed in the function. Ex. 9798 = 9.798 m/s
2
Step 1 Press the CAL key in normal weighing mode. The
Cal in is displayed for two seconds. Press the
M+ key to enter the gravity compensation function.
If you want to cancel the current procedure, press
and hold the
NET/GROSS
key and press the CAL key.
Then, all parameters are not changed and the
calibration mode is finished. Press the
Operate/Standby
key
to turn the display off after displaying CanCel .
Step 2 The parameter g1 xxxx is displayed. Enter the
gravity acceleration of the calibration location. The
parameter xxxx is the gravity acceleration.
key...........The key to select a figure.
key...........The key to select a number.
key...........The key to store the new gravity
acceleration and proceed to step 3.
MODE key ......The key not to change the value
and proceed to step 3.
Step 3 The parameter g2 xxxx is displayed. Enter the
gravity acceleration of the installed location. The
parameter xxxx is the gravity acceleration.
key...........The key to select a figure.
key...........The key to select a number.
key...........The key to store the new gravity
acceleration and proceed to step 4.
MODE key ......The key not to change the value
and proceed to step 4.
Step 4 Now g1 xxxx is displayed. Press the CAL key to
store the parameters. The end is displayed.
Proceed to step5.
Step 5 Press the
Operate/Standby
key to turn display off.
5.5. Gravity Compensation Function