Product specifications
AD-4329A Weighing Indicator Page 21
Step 6 Specify the weighing range.
MODE key ..The key to use a single weighing
range and proceed to step 10.
key....... The key to store a value as dual
range and proceed to step 10.
PRESET TARE
key. The key to store the second range
and proceed to step 7.
Setting Third Range
Step 7 The
mark is displayed at net and gross.
Select a way.
Case 1
If keeping the current third division, wait for
displaying Cxxxx and proceed to step 8.
The xxxx is the range.
Case 2
If changing the third division, press the key
while displaying Cap3-xx . The xx is division.
key....... The key to select the division.
key....... The key to store the new division
and proceed to step 8.
MODE key .. The key not to change the division
and proceed to step 8.
Step 8 Specify the range of the third range.
key....... The key to select a figure.
key....... The key to select a number.
Step 9 Specify the weighing range.
MODE key ..The key to use dual weighing
range and proceed to step10.
key....... The key to store the value as
triple range and proceed to
step10.
Step10 Cal 0 is displayed. Perform zero calibration
and span calibration from step7 at page 15 .
5.3. Weighing Range Function