Product specifications
AD-4329A Weighing Indicator Page 19
5.3.1. Setting Division and Range
Consider the following rules to design the weighing range.
Rule 1 Select the division and range of each weighing range so as to fit the following inequality.
The first range < the second range < the third range
The division of next weighing range is automatically set bigger than division of
lower weighing range. And the division can change.
Rule 2 When setting the dual range, the upper limit value of the second range becomes
the maximum capacity.
When setting the triple weighing range, the upper limit value of the third range
becomes the maximum capacity.
Rule 3 Select a resolution smaller than 10000. The resolution is a value divided the
maximum capacity by the minimum division of the first range.
Mark to indicate the weighing range while this setting
mark First range Second range Third range
Net ON
Gross OFF
Procedure
Step 1 Press the CAL key to enter the calibration mode.
Then Cal in is displayed for two seconds.
MODE key .. The parameters are not revised and the next
item is displayed.
CAL key ..... Parameters are stored in the indicator and
Eend is displayed. Press the
Operate/Standby
key to turn the display off.
Press and hold the
NET/GROSS
key and press the CAL key
All parameters are not changed and the
calibration mode is finished. Press the
Operate/Standby
key to turn the display off
after displaying CanCel .
5.3. Weighing Range Function