Product specifications
AD-4329A Weighing Indicator Page 17
Digital Input
Step16 Pressing the M+ key, a stored input voltage
parameter of the span is displayed in the unit
of mV/V. Adjust the input voltage using the
following keys.
key ........ The key to select a figure.
key ........ The key to select a number.
key ........ The key to store the span parameter
and proceed to step 17.
MODE key .. The key not to change parameter
and go back to step 4.
CAL key...... The key not to change parameter
and proceed to step 17.
5.2.8. Exiting the Calibration Mode
Step17 Check the Cal end display.
Use the following keys.
CAL key ..... Parameters are stored in the indicator
and end is displayed. Proceed to
step 18.
MODE key .. The key to memorize parameters
temporarily. Proceed to the "Minimum
Division Selection".
Press and hold the
NET/GROSS
key and press the CAL key
All parameters are not changed,
CanCel is displayed and the
calibration mode is finished.
Step18 Press the
Operate/Standby
key to turn the display off.
5. Calibration
From last page, B From last page, C
From step10
From step1
3
From step1
5
Stored in
p
ut
v
olta
ge
New input voltage
mV
/
V
mV
/
V
Turn display off