Product specifications

Page 16 AD-4329A Weighing Indicator
5.2.7. Span Calibration
Step10 Check the fxxxx display. The xxxx is a capacity.
Select a span calibration method to adjust the capacity.
Weighing a mass except the
maximum capacity
The method to weigh a mass except
the maximum capacity.
To step 11
Weighing maximum capacity
The method to weigh a mass
equivalent to the maximum capacity.
To step 14
Digital input
The numerical input to enter a load
cell output voltage.
To step 16
Weighing a Mass except the Maximum Capacity
Step11 Set a mass value using the follwing keys.
key .... Key to select a figure.
key .... Key to select a number.
Step12 Place a mass equivalent to displayed value on the
weighing unit.
Step13 Press the key after the MD mark turns off.
Proceed to step 17.
MODE key .... The key not to change parameter
and go back to step 4.
CAL key.......The key not to change parameter
and proceed to step 17.
Caution Do not press the key while the MD mark
is lit (detecting motion). Arrange the condition
using the F00 filter function.
Weighing Capacity Mass
Step14 Place a mass equivalent to the maximum capacity on
weighing unit.
Step15 Press the key after the MD mark turns off.
Proceed to step 17.
MODE key .... The key not to change parameter
and go back to step 4.
CAL key.......The key not to change parameter
and proceed to step 17.
Caution Do not press the key while the MD mark
is lit (detecting motion). Arrange the condition
using the F00 filter function.
5. Calibration