Product specifications
Page 14 AD-4329A Weighing Indicator
5.2.4. Minimum Division (Weighing Interval)
Step 4 Check the d-xx display. The xx is minimum division (01, 02, 05, 10, 20, 50).
Step 5 Select a minimum division using the following keys.
key....... The key to select a parameter.
key....... The key to store a parameter and
proceed to step 6.
MODE key... The key not to change a parameter
and proceed to step 6.
M+ key...... The key to proceed to "5.5. Gravity
Compensation Function".
5.2.5. Maximum Capacity
Step 6 Check the Cxxxx display. The xxxx is current
capacity. Select a new maximum capacity
using the following keys.
key...........The key to select a figure.
key...........The key to select a number.
key...........The key to store a parameter
and proceed to step 7 (of the
zero calibration).
MODE key......The key not to change a
parameter and proceed to step
7 (of the zero calibration).
PRESET TARE
key .... The key to porceed to "5.3.
Weighing Range Function".
5. Calibration