Product specifications
AD-4329A Weighing Indicator Page 13
Step H Gravity compensation function.
Perform this function after span calibration.
5.2.2. Preparation
Step 1 Keep the following conditions to calibrate the scale (indicator) correctly.
Maintain a constant temperature, stable power and stable input voltage from the
load cell.
Avoid direct sunshine and in the neighborhood of an air conditioner.
Do not install the scale (indicator) where there is a strong magnetic field.
Step 2 Turn the display on during 10 minutes.
5.2.3. Entering the Calibration Mode
Step 3 Press the CAL key in the normal weighing mode. The Cal in is displayed for
two seconds when entering the calibration mode.
MODE key .. Parameters are not revised and the next item is displayed.
CAL key ..... Parameters are stored in the indicator and end is displayed.
Press the
Operate/Standby
key to turn the display off.
Press and hold the
NET/GROSS
key and press the CAL key
.................... All parameters are not changed
and the calibration mode is
finished. Press the
Operate/Standby
key to turn the display off after
displaying CanCel .
5. Calibration