Maintenance
Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you
cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our
customer service department. The A&D customer service will provide technical information, spare parts and units
to authorized dealers.
The device was designed and manufactured for a long service life. However it is generally recommended to have
the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact the authorized
dealer in your area or A&D for maintenance.
Technical Data
Type
UB-542
Clinical test
According to ANSI / AAMI SP-10 1992
EMC IEC 60601-1-2: 2007 Measurement
method
Oscillometric measurement
Memory Last 90 measurements
Measurement
range
Pressure: 0 – 299 mmHg
Pulse:
40 – 180 beats / minute
Operating
condition
+10°C to +40°C / 30%RH to 85 %RH
Measurement
accuracy
Pressure: ±3 mmHg
Pulse: ±5%
Storage
condition
-20°C to +60°C / 10%RH to 95 %RH
Power supply
2 x 1.5V alkaline batteries (LR03 or AAA)
Dimensions
Approx. 56 [W] x 88 [H] x 18 [D] mm
Wrist
circumference
13.5 – 21.5 cm
Weight
Approx. 103 g, excepting batteries
Classification
Type BF
Note: Specifications are subject to change for improvement without prior notice.
EMC table information is listed on our website: http://www.aandd.jp/products/manual/medical/emc_en.pdf