Wrist Digital Blood Pressure Monitor Model UB-542 Instruction Manual Original Manuel d’instructions Manual de Instrucciones Manuale di Istruzioni 使用手冊 1WMPD4002391A
Contents Dear Customers..................................................................................................................................2 Preliminary Remarks ...........................................................................................................................2 Precautions.........................................................................................................................................2 Parts Identification..........................................
Dear Customers Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using the device for the first time. Preliminary Remarks This device conforms to the European Directive 93/42 EEC for Medical Products. This is made evident by the mark of conformity.
Parts Identification button SET button Display Case START button Battery cover Cuff Alkaline batteries (LR03 or AAA) Display Systolic pressure WHO classification indicator and Pressure bar indicator Diastolic pressure Heart mark Pulse rate MEMORY Average Battery indicator AM / PM mark I.H.B.
Symbols Symbols that are printed on the device case Symbols Function / Meaning Standby and Turn the device on Battery installation guide Direct current SN Serial number Date of manufacture Type BF: Device and cuff are designed to provide special protection against electrical shocks.
Symbols Symbols that appear on the display (continued) Symbols Function / Meaning Previous measurements stored in MEMORY. Average data Full Battery The battery power indicator during measurement. Low Battery The battery is low when it blinks. or PUL. DISPLAY ERROR Unstable blood pressure due to movement during measurement. The systolic and diastolic values are within 10 mmHg of each other. The pressure value did not increase during inflation. The cuff is not applied correctly. DIA. PUL./min.
Using the Monitor 1. Installing / Changing the Batteries 1. Remove the battery cover. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) are correct. Use only LR03 or AAA batteries. 3. Attach the battery cover. • Insert the batteries as shown in the battery compartment. If installed incorrectly, the device will not work. • When (LOW BATTERY mark) appears on the display, replace all batteries with new ones.
Using the Monitor 2. Adjusting the Built-in Clock Press Adjust the clock prior to use. 1. Press the SET button until the year starts blinking. 2. Select the year using the Year button. Press the SET button to set the current year and move to month/day selection. The date can be set anywhere between the years 2011 and 2059. Month 3. Select the month using the button. Press the SET button to set the current month and move to day selection. 4. Select the day using the button.
Using the Monitor 3. Applying the Cuff 1. Wrap the cuff around your wrist about 1 cm above your hand as shown in the figure at the right. 2. Fasten the cuff tightly using the Velcro strip. Note: For proper measurements, fasten the cuff tightly and measure on a bare wrist. About 1 cm 4. How to Take Accurate Measurements For the most accurate blood pressure measurement: • Sit comfortably at a table. Rest your arm on the table. • Relax for about five to ten minutes before measurement.
Measurements Note: The UB-542, once used, will provide an inflation appropriate to the user, based on the previous measurement. 1. Sit comfortably with the cuff at the same level as your heart and relax. 2. Press the START button. All of the display segments are displayed. Zero (0) is displayed blinking briefly. Then the display changes, as indicated in the figure at the right, as measurement begins. The cuff starts to inflate. It is normal for the cuff to feel very tight.
Measurements Notes For Accurate Measurement Sit down in a comfortable position. Place your arm on a table or other support. Raise your hand so that the cuff is at the same level as your heart. Relax for about five to ten minutes before taking a measurement. If you are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.
Recalling the Memory Data Note: This device stores the last 90 measurements in memory. Average of 1. Press the button. all the data The average of all measurements and the number of data are displayed. (If no data, “0” is displayed. Press the or START button to turn the device off.) 2. Each time the button is pressed, the memory data is displayed as follows. Average of all AM (morning) measurements taken between 4:00 and 9:59. (In the example, 8 measurements. If no data, “--” is displayed.
What is an Irregular Heartbeat? Model UB-542 blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you be relaxed, remain still and do not talk during measurements. Note: We recommend contacting your physician if you see this symbol frequently.
WHO Classification Indicator Each six segments of the bar indicator correspond to the WHO blood pressure classification described on the next page. Example WHO Classification Indicator Severe hypertension Moderate Mild High hypertension hypertension normal Moderate hypertension Mild hypertension High normal Normal Optimal : The indicator displays a segment, based on the current data, corresponding to the WHO classification.
To prevent hypertension or keep it under control: Do not smoke Exercise regularly Reduce salt and fat intake Have regular physical checkups Maintain proper weight Why Measure Blood Pressure at Home? Blood pressure measured at a clinic or doctor's office may cause apprehension and can produce an elevated reading, 25 to 30 mmHg higher than that measured at home.
Troubleshooting Problem Possible Reason Batteries are drained. Nothing appears in the display, even when the Battery terminals are not in the correct device is turned on. position. The cuff does not inflate. Battery voltage is too low. (LOW BATTERY mark) blinks. If the batteries are drained completely, the mark does not appear. The cuff is not applied properly. You moved your arm or body during the measurement. Recommended Action Replace all batteries with new ones.
Maintenance Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The A&D customer service will provide technical information, spare parts and units to authorized dealers. The device was designed and manufactured for a long service life.
1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LTD. Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131 U.S.A. Telephone: [1] (408) 263-5333 Fax: [1] (408)263-0119 A&D Australasia Pty Ltd.