Instruction manual

English 8
Using The Monitor
Alarm ON/OFF Mark On The Display
When the alarm is on, a “ ” appears on the LCD. To turn the alarm off,
simply press the Clock button once.
The alarm will sound for 1 minute. To stop the alarm sound, press any
button once.
Selecting The Correct Cuff Size
Using the correct cuff size is important for an accurate reading. If the cuff is
not the proper size, the reading may yield an incorrect blood pressure value.
The arm size is printed on each cuff.
The arm cuff is a consumable. If it becomes worn, purchase a new one.
Arm Size Recommended Cuff Size Catalog Number
22 cm to 43 cm Large adult cuff
CUF-U-LA
17 cm to 32 cm Adult cuff
CUF-U-A
Arm size: The circumference of the biceps.
Note: Model UA-855 is not designed for using a small adult cuff.
Symbols that are printed on the cuff
Symbols Function/Meaning Recommended Action
Artery Position Mark
Set the mark on the artery of the
upper arm or in line with the ring finger on
the inside of the arm.
REF Catalog Number
Lot Number
Applying The Arm Cuff
1. Wrap the cuff around the upper
arm, about 1-2 cm above the
inside of the elbow, as shown.
Place the cuff directly against the
skin, as clothing may cause a faint
pulse, and result in a measurement
error.
2. Constriction of the upper arm,
caused by tightly rolling up a
shirtsleeve, may prevent accurate
readings.
Do not roll up shirtsleeve tightl
y
1 to 2 cm
A
ir hose
Cuff
A
rtery position mark
Secure the fabric
fastener firmly