Instruction manual
English 5
Symbols
Symbols Function / Meaning Treatment
Turns on or off the device
Sets the clock / Does not store the current data
Direction guide to install battery
Direct current
SN
Serial number
Date of manufacture
Type BF: Device, cuff and tubing are
designed to provide special protection
against electric shocks.
Appears when measurement is in progress
and flashes when the pulse is detected.
Measurement is in progress --
remain as still as possible.
Previous measurement stored in MEMORY
A
ppears when the battery voltage is
too low for the device to work properly.
Replace all batteries with new
ones.
Appears if the systolic and diastolic
measurements are within 10 mmHg of
each other.
Verify that the air hose is
properly connected and that the
exhaust rate is between 2 and 5
mmHg. Check for air leakage.
A
ppears if the pressure value is
unstable due to movement during the
measurement.
Try the measurement again.
Remain very still during the
measurement.
Appears if the pressure value did not
increase during pump up.
Verify that the air hose is
properly connected and that the
exhaust rate is between 2 and 5
mmHg. Check for air leakage.
System error
Remove the batteries. Install
them again properly.
A
ppears if pulses are not detected
correctly.
Apply the cuff properly and try
the measurement again.
Attention symbol See instruction for use.
SYS Systolic blood pressure in mmHg
DIA Diastolic blood pressure in mmHg
PUL.
/min
Pulse per minute
EC directive medical device label
WEEE label
Manufacturer
REPEC
EU-representative
Refer to instruction manual/booklet
Class II device
Polarity of DC jack