Do not open the device. It uses delicate electrical components and an intricate air
unit that could be damaged. If you cannot fix the problem using the troubleshooting
instructions, contact the authorized dealer in your area or our customer service
department. The A&D customer service will provide technical information, spare
parts and units to authorized dealers.
The device was designed and manufactured for a long service life. However it is
generally recommended to have the device inspected every 2 years, to ensure
proper functioning and accuracy. Please contact the authorized dealer in your
area or A&D for maintenance.
Technical Data
Type UA-767NFC
Measurement method Oscillometric measurement
Measurement range Pressure: 0 - 280 mmHg
Pulse: 40 - 180 beats / minute
Measurement accuracy Pressure: ±3 mmHg
Pulse: ±5%
Power supply 4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233) (Not included)
Number of measurements Approximately 450 measurements, when AA
Alkaline batteries are used, with pressure value
of 180 mmHg at room temperature of 23 °C.
Classification Internally powered ME equipment (Supplied by
batteries) /
Class II (Supplied by adapter)
Continuous operation mode
Clinical test According to ANSI / AAMI SP-10 1992
EMC IEC 60601-1-2: 2007
Memory 0 or 100 measurements (Only for devices that
rewriting of the memory data is enabled.)
Operating condition +10 °C to +40 °C / 15 %RH to 85 %RH /
800 hPa to 1060 hPa
Transport / Storage condition -10 °C to +60 °C / 15 %RH to 95 %RH
Dimensions Approx. 147 [W] x 64 [H] x 110 [D] mm
Weight Approx. 300 g, excluding the batteries
Applied part Cuff Type BF