Instruction manual

English 15
Press to increase the memory number by
one.
Press to decrease the memory number by
one.
Press to return to the clock display.
Press and
hold
1.5 seconds
1.5 seconds
1.5 seconds
Clock displa
y
Memory number
Date Time
START
START
Near Field Communication (NFC) Function
To transmit data to or from the device, communication devices that comply with
the communication specifications of the device are required.
Examples of communication devices that comply
NFC-enabled cellular phone
Personal computer with NFC reader/writer
Access Point
Displaying the Memory Data
When the memory setting is 100, the memory data can be displayed.
While the device displays the clock, press and hold the START button.
The systolic pressure, diastolic pressure and pulse rate with the memory number
are displayed. At every 1.5 seconds, the memory number display changes to the
date, then to the time, and then to the memory number again.
Press the
button to increase the memory number by one and press the
button to decrease the memory number by one.
Press the
START button to return to the clock display. After one minute of
non-operation, the device will return to the clock display.
Note: Displaying the memory data is not available when