Digital Blood Pressure Monitor Model UA-767NFC Instruction Manual Original Manuel d’instructions Traduction Manual de Instrucciones Traducción Manuale di Istruzioni Traduzione Bedienungsanleitung Übersetzung 1WMPD4002380D
Contents Dear Customers ............................................................................................ 1 Preliminary Remarks...................................................................................... 2 Precautions.................................................................................................... 2 Parts Identification......................................................................................... 3 Symbols....................................................
Preliminary Remarks This device conforms to the European Directive 93/42 EEC for Medical Products. mark of conformity. This is made evident by the (0123: The reference number to the involved notified body) The device is designed for use on adults only, not newborns or infants. Environment for use. The device is for use in the home healthcare environment. This device is designed to measure blood pressure and pulse rate of people for diagnosis. Precautions Precision components are used in this device.
Do not use the device where flammable gases such as anesthetic gases are present. It may cause an explosion. Do not use the device in highly concentrated oxygen environments, such as a high-pressure oxygen chamber or an oxygen tent. It may cause a fire or explosion. Parts Identification Air Hose Arm Cuff Proper Fit Range DC Jack Communication Area Index Mark Artery Position Mark Air Connector Plug START Button Air Socket Display Battery Compartment Clock Button Button Battery Cover 1.
Symbols Symbols that are printed on the device case Symbols SYS. DIA. PUL./min. 2010 SN Function / Meaning Standby and Turn the device on. Systolic blood pressure in mmHg Diastolic blood pressure in mmHg Pulse per minute Battery installation guide Direct current Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks.
Symbols Symbols Full Battery Low Battery Function / Meaning The battery power indicator during measurement. The battery is low when it blinks. Unstable blood pressure due to movement during measurement. The systolic and diastolic values are within 10 mmHg of each other. The pressure value did not increase during the inflation. The cuff is not applied correctly. PUL. DISPLAY ERROR Recommended Action Replace all batteries with new ones when the mark blinks. Take another measurement.
Operation Mode 1. Normal Measurement Press the START button. Blood pressure is measured. 2. Memory Function When the memory setting is 100: This device will store the last 100 measurements in memory. When the measurement exceeds 100, the oldest data will be deleted automatically and the most recent data will be stored. When the memory setting is 0: This device will not store any data in memory. 3. Deleting all Data Stored in Memory Make sure that the device displays the clock. button.
Using the Monitor Installing / Changing the Batteries Step 1 1. Slide the battery cover up to open it. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+ and -) are correct. Use only R6P, LR6 or AA batteries. 3. Slide the battery cover down to close. Step 3 Step 2 Connecting the AC Adapter Insert the AC adapter plug into the DC jack. Next, connect the AC adapter to an electrical outlet. Use the specified AC adapter.
Using the Monitor Connecting the Air Hose Insert the air connector plug into the air socket firmly. Air socket Air connector plug Setting and Adjusting the Clock Set or adjust the clock prior to use. Press and hold 1. Press and hold the button until the year starts blinking. “Year” blinks when inserting the batteries or connecting the AC adapter Year When inserting the batteries or connecting the AC adapter, “Year” blinks automatically. button. 2.
Notes on the Built-in Clock The device displays the clock as long as power is connected. Adjust the clock prior to use. During measurement, the clock adjustment is not available. button to increase the value by one, or When adjusting the clock, press the button to change the value continuously. press and hold the The setting procedure cannot be reversed. If the wrong value is set, start the setting procedure again.
Using the Monitor Selecting the Correct Cuff Size Using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit range, on the cuff, tell you if you are applying the correct cuff. (Refer to "Symbols that are printed on the cuff" below) If the index points outside of the range, contact your local dealer to purchase a replacement cuff.
Using the Monitor Applying the Arm Cuff 1. Wrap the cuff around the upper arm, about 1 - 2 cm above the inside of the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse, and result in a measurement error. Do not roll up shirtsleeve tightly Cuff 1 to 2 cm 2. Constriction of the upper arm, caused by tightly rolling up a shirtsleeve, may prevent accurate readings. Artery position mark Secure the fabric fastener firmly points 3.
Measurements Before measurement, read “Notes for Proper Measurement” on page 14. Normal Measurement 1. Place the cuff on the arm (preferably the left arm). Sit quietly during measurement. 2. Press the START button. All of the display segments are displayed. Zero (0) is displayed blinking briefly. The display changes, as indicated in the figure at the right, as the measurement begins. The cuff starts to inflate. It is normal for the cuff to feel very tight.
Measurements Measurement with the Desired Systolic Pressure Model UA-767NFC is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. Use this method when re-inflation occurs repeatedly or when the results are not displayed even if the At heart level pressure decreases to 20 mmHg or less. When the memory setting is 0: 1. Place the cuff on the arm (preferably the left arm). 2.
Notes for Proper Measurement Sit down in a comfortable position. Place your arm on a table with your palm facing upward and the cuff at the same level as your heart. Relax for about five to ten minutes before taking a measurement. If you are excited or depressed by emotional stress, the measurement will reflect this stress as a higher (or lower) than normal blood pressure reading and the pulse reading will usually be faster than normal.
Near Field Communication (NFC) Function To transmit data to or from the device, communication devices that comply with the communication specifications of the device are required. Examples of communication devices that comply NFC-enabled cellular phone Personal computer with NFC reader/writer Access Point Displaying the Memory Data When the memory setting is 100, the memory data can be displayed. While the device displays the clock, press and hold the START button.
Switching the Memory Setting While the device displays the clock, hold the communication symbol of the NFC-enabled communication device close to the communication area ( ) of the ” appears. The memory setting is switched. device until “ Initiate communication Data transmission 100 blinks. Clock display Transmission starts Initiate communication Transmission ends (The memory setting has been switched to 100.) Data transmission 0 blinks.
Transmitting the Memory Data While the device displays the clock, hold the communication symbol of the NFC-enabled communication device close to the communication area ( ) of the ” appears. The memory data transmission is performed. If the device until “ clock of the device is not correct, the clock will be updated at the same time.
What is an Irregular Heartbeat The UA-767NFC provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you are relaxed, remain still and do not talk during measurements. Note: We recommend contacting your physician if you see this frequently.
About Blood Pressure What is Hypertension and How is it Controlled? Hypertension, an abnormally high arterial blood pressure, if left unattended, can cause many health problems including stroke and heart attack. Hypertension can be controlled by altering lifestyle, avoiding stress, and with medication under a doctor’s supervision.
Troubleshooting Problem Possible Reason Recommended Action Replace all batteries with new ones. Nothing appears Batteries are drained. on the display, Reinstall the batteries with negative even when the Battery terminals are not in and positive terminals matching power is turned the correct position. those indicated on the battery on. compartment. Battery voltage is too low. (LOW BATTERY mark) The cuff does not Replace all batteries with new ones. blinks. If the batteries are inflate.
Maintenance Do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, contact the authorized dealer in your area or our customer service department. The A&D customer service will provide technical information, spare parts and units to authorized dealers. The device was designed and manufactured for a long service life.
Useful life Device: Cuff: 5 years (when used six times a day) 2 years (when used six times a day) Wireless communication NFC wireless interface module Accessory AC adapter The adapter is to connect the device to a power source at home. TB-233 Please contact your local A&D dealer for purchasing. The AC adapter is required to be inspected or replaced periodically.
1-243 Asahi , Kitamoto-shi, Saitama 364-8585, JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LIMITED Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D ENGINEERING, INC. 1756 Automation Parkway, San Jose, California 95131, U.S.A.