Instruction manual

English 23
Maintenance
Do not open the device. It uses delicate electrical components and an
intricate air unit that could be damaged. If you cannot fix the problem using
the troubleshooting instructions, contact the authorized dealer in your area
or our customer service department. The A&D customer service will
provide technical information, spare parts and units to authorized dealers.
The device was designed and manufactured for a long service life.
However it is generally recommended to have the device inspected every
2 years, to ensure proper functioning and accuracy. Please contact the
authorized dealer in your area or A&D for maintenance.
Technical Data
Type UA-767F
Measurement method Oscillometric measurement
Measurement range Pressure: 0 - 299 mmHg
Systolic pressure: 60 - 279 mmHg
Diastolic pressure: 40 - 200 mmHg
Pulse: 40 - 180 beats / minute
Measurement accuracy Pressure: ±3 mmHg
Pulse: ±5%
Power supply 4 x 1.5V batteries (R6P, LR6 or AA) or
AC adapter (TB-233) (Not included)
Number of measurements
Approx. 700 times LR6 (
alkaline batteries
)
Approx. 200 times R6P (
manganese batteries
)
With pressure value 180 mmHg, room temperature
23 °C.
Classification Internally powered ME equipment (Supplied by
batteries) / Class II (Supplied by adapter)
Continuous operation mode
Clinical test According to ANSI / AAMI SP-10 1992
EMC IEC 60601-1-2: 2007
Memory Last 60 measurements for each of the four users
Operating conditions +10 to +40 °C / 15 to 85 %RH / 800 to 1060 hPa
Transport / Storage conditions
-20°C to +60°C / 10%RH to 95%RH
Dimensions Approx. 140 [W] x 60 [H] x 105 [D] mm
Weight Approx. 255 g, excluding the batteries
Ingress protection Device: IP21