Specifications
 18
8. COMMUNICATION SPECIFICATIONS 
The TM-2655/TM-2655P is equipped with two RS-232C channels. Various settings for each channel are 
available in the function setting mode. Refer to “7-2 Description of the Functions” for details. 
8-1  Channel 1 : Miniature DIN 8-pin 
 CAUTION 
  The personal computer and medical equipment connected to the device are not allowed to 
be in the patient area. 
  The personal computer used must conform to IEC60950 
Specifications 
Table 6 Channel 1 specifications 
Standard  Conforms to EIA RS-232C 
Transmission Asynchronous, half-duplex 
Baud rate  1200, 2400, 4800, 9600 bps (Can be changed in the function setting mode “F07”.)
Data bits  8 bits 
Parity bit  None 
Stop bits  1 bit 
Code ASCII 
Equipment available for connection 
Channel 1:  Scales, and automatic weight and height scales manufactured by A&D   
Channel 2:  Personal computer 
Pin assignment 
Pin No.  Signal name Description 
1 TXD Send data 
2 RXD Receive data 
3  RTS  Ready to send 
4 - Not connected 
5  CTS  Clear to send 
6 GND Signal ground 
7 - Not connected 
8 - Not connected 
Fig.14 Pin assignment 
Cable connection between the device and a personal computer 
   TM-2655/TM-2655P  Personal computer 
Fig.15 Cable connection diagram 
Note 
Do not use pins 4, 7, and 8. 
They are used by the device.