7-6 Error Codes 
When the instrument detects an erroneous measurement condition, the following error codes are 
displayed. 
Error code  Symptom  Remedy 
Blood pressure related errors 
E00 
Zero point error in the pressure 
sensor circuit 
Exhaust air from the cuff and turn on the power 
again. 
E11  Can not pressurize. 
E12  Pressurizing speed is too slow. 
Check the cuff and hose for correct connection or 
check them for being folded. 
E13  Pressurizing speed is too fast. 
Confirm that the patient setting is correct. (Adult, 
Neonate) 
E21 
Measurement time is too long. 
Constant exhaustion speed is 
too slow. 
E22  Exhaustion speed is too fast. 
E23  Excessive pressure is detected. 
Check the cuff and hose for correct connection or 
check them for being folded. 
E42 Insufficient pressurization 
E43  No pulse is detected. 
E44 Physical movement is detected. 
E45 
The diastolic blood pressure 
cannot be determined. 
E46 
The mean blood pressure 
cannot be determined. 
E48 
The systolic blood pressure 
cannot be determined 
E61 
The pulse rate cannot be 
determined. 
E63 
The blood pressure value is 
inappropriate. 
Check the cuff for correct placement or check the 
patient for physical movement or for irregular pulse.
SpO
2
 related errors 
E30 SpO
2
 self test error  Turn on the power again. 
E31 SpO
2
 sensor error  Confirm that the specified sensor is connected. 
ECG related errors 
E35  ECG self test error 
E36  Defective ECG circuit 
Turn on the power again. 
Other error 
E80  System error  Turn on the power again. 
Note 
•  Cuffs and sensors are consumables. If a measurement error occurs frequently, they must be replaced. 
Refer to "8 Accessories/Options" for detail information. 